Tuesday, September 26, 2017

Amphetamines confirmation (with chromatogram), urine

Reference values Modification

Test code: 5255

Effective update from 29/09/2017




PREVIOUS

NEW
        
      Negative: Less than 200 ng/mL of amphetamine, methamphetamine, MDA and MDMA. EWDTS 2011 (European Workplace Drug Testing Society).
     
      *LOD: 10 ng/mL of amphetamine, methamphetamine, MDA and MDMA.
      * Fonte: Bula do fabricante



Amphetamine:
Quantification limit : 25   ng/mL
Measuring uncertainty:  6,1 ng/mL
     
Methamphetamine:
Quantification limit : 25   ng/mL
Measuring uncertainty:  5,4 ng/mL

MDA (methylenedioxyamphetamine):
Quantification limit : 25   ng/mL
Measuring uncertainty:  3,8 ng/mL

MDMA (methylenedioxymethamphetamine):
Quantification limit : 25   ng/mL
Measuring uncertainty:  4,5 ng/mL
     
In the received sample all of the indicated substances have been analyzed. The following positivity criterion has been used: when the concentration of the substances is over or equal the quantification limit value plus the value of test uncertainty, the result is Positive. In opposite case the result is Negative. The results obtained are only related to the sample tested.


The laboratory has a measurement uncertainty across the entire working range.  
                
         
         
         
         

   Find the record of the test by clicking here

No comments:

Post a Comment